← All products
A12/00779/21

Particle filter mask

KN95 mask

Protective equipmentHealth risk / other
Notifying country
Iceland
Sold online
No
Open the alert on the European Commission website ↗

What is the danger

The particle/filter retention of the material is insufficient (measured values: as low as 73%) and the total filtration capacity of the mask is insufficient (measured mean values: as low as 60%). Furthermore, the mask does not fit correctly to the face of the user. Consequently, an excessive amount of particles or microorganisms might pass through the mask; thus, even if combined with other recommended measures, it may not properly protect.

Requirements breached

The product does not comply with the Personal Protective Equipment Regulation and with the relevant European standard EN 149.

Measures taken

  1. Recall of the product from end users Voluntary measuresTaken by: DistributorIn force from: 23 March 2021

Product

Description
White respiratory protective half-face mask.
Packaging
The product is sold in a white carboard box containing 10 pieces.

Original source

Data comes from the official Safety Gate system of the European Commission.

Alert number
A12/00779/21 ↗
Weekly report
Report-2021-22
Published
3 June 2021
Modified
3 June 2021